🌡️ Thermometer Verification

PKKK/200/009 — Verifikasi Termometer

Procedure for verifying test thermometers against calibrated master thermometers. Ensures temperature measurement accuracy across the laboratory.

Thermometer Verification Procedure
◈ 1. Preparation & Environmental Equilibrium

Essential requirements before initiating thermometer verification to ensure valid comparative readings between Master and Test instruments.

◈ 2. Verification Procedure (PKKK/200/009)
3 Mo
Frequency
5
Readings Required
≤1.0°C
Tolerance

Systematic recording of comparative temperatures ensuring proper intervals and stabilization.

  1. Verify the physical condition of the test thermometer (no separated liquid columns, no cracks, clear scale).
  2. Check the calibration certificate status of the Master Thermometer and ensure the required range is covered.
  3. Immerse both thermometers closely together in the medium. Ensure bulbs/sensors do not touch the walls or bottom of the container.
  4. Allow at least 1 hour for stabilization.
  5. Record the first paired reading (Master $T_{M1}$ and Test $T_{T1}$).
  6. Wait for an interval of at least 5 minutes.
  7. Record the second paired reading ($T_{M2}$, $T_{T2}$).
  8. Repeat the 5-minute intervals until a total of 5 consecutive paired readings have been recorded.
  9. Record details of the Master Thermometer in the report including: Resolution, Calibration Date, Calibration Range, and Calibration Certificate Number.
◈ 3. Calculations & Linearity Equation

The true value of the Test Thermometer ($y$) must be calculated using the Master Thermometer's calibration linearity equation.

Linearity Equation from Master Certificate:
y = mx + c

Where:
$y$ = True Temperature (°C) derived from Master
$m$ = Gradient (Slope) from calibration cert
$x$ = Actual reading shown on Master Thermometer
$c$ = y-intercept from calibration cert
◈ 4. Acceptance Criteria & Troubleshooting
✅ Acceptance Criteria

The maximum temperature difference between the derived Master true value ($y$) and the Test thermometer reading ($T_{Test}$) shall not exceed $1.0^{\circ}\text{C}$ for all 5 readings.
Tolerance limit: $\le 1.0^{\circ}\text{C}$.

⚠️ Actions on Failure

If the Test thermometer fails to meet the $\le 1.0^{\circ}\text{C}$ specification:
1. Do not immediately discard the thermometer.
2. Repeat the entire verification process by a different analyst (Second checking).
3. If the second verification also fails, the test thermometer is deemed 'Out of Tolerance' and must be removed from use or sent for official recalibration/repair.

📋 Reference Forms

PKK/047A — Thermometer Verification Record Sheet
PKK/047B — Generation of Linearity Equation for Actual Temperature Reading